Repeatability and reproducibility of the galilei G6 and its agreement with the pentacam® AXL in optical biometry and corneal tomography
Version 1To assess the repeatability and reproducibility of corneal measurements performed by the Galilei G6 and determine its agreement with the Pentacam® AXL. 105 eyes underwent examination by both devices, measuring nine parameters. Paired t-tests, Deming Regression, and Bland-Altman plots were used to determine agreement. Analysis of Variance was used to determine repeatability and reproducibility. Measurements showed no clinically significant differences between the two devices. On average, the Gali
No license available
To assess the repeatability and reproducibility of corneal measurements performed by the Galilei G6 and determine its agreement with the Pentacam® AXL. 105 eyes underwent examination by both devices, measuring nine parameters. Paired t-tests, Deming Regression, and Bland-Altman plots were used to determine agreement. Analysis of Variance was used to determine repeatability and reproducibility. Measurements showed no clinically significant differences between the two devices. On average, the Gali