Repeatability and Reproducibility of Slit Lamp, Optical Coherence Tomography, and Scheimpflug Measurements of Corneal Scars
Version 1Purpose: To determine the repeatability and reproducibility of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug photography for several measurements of corneal scars, including scar size, scar depth, and corneal thickness. Methods: A series of patients treated for fungal keratitis at a tertiary eye care center in South India were recalled two years after successful treatment. Eyes with corneal scars had a slit lamp examination performed by two ophthalmologists masked to the
No license available
Purpose: To determine the repeatability and reproducibility of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug photography for several measurements of corneal scars, including scar size, scar depth, and corneal thickness. Methods: A series of patients treated for fungal keratitis at a tertiary eye care center in South India were recalled two years after successful treatment. Eyes with corneal scars had a slit lamp examination performed by two ophthalmologists masked to the